GL7-L/P
This is a digital input module capable of measuring the timing of the
control signals or the number of pulses.
control signals or the number of pulses.
In Logic mode: sampling in 1MS/s (1μs) to 1S/hr. (1hr.).
It is able to measure the timing of the control signals as the relay
by this module. It allows to measure simultaneously both the timing
of control signals and an analog waveform using other analog input modules.
In Pulse mode: sampling in 10kS/s (100μs) to 1S/hr.(1hr.) It is able to count
the number of pulses for measuring such as the encoder output, a rotation
speed, or flow rate. It allows to measure simultaneously both the number of
pulses and an analog waveform using other analog input modules.
Application examples
GL7-L/P specification
Item | Description | |
---|---|---|
Type of module | Logic or Pulse measurement module | |
Model number | GL7-L/P | |
Number of input channels | 16 channels | |
Input method |
All channels common ground, Simultaneous sampling, Circular connector (4ch/connector) (*1) |
|
Sampling speed (interval) | Logic mode | 1 M Samples/s to 1 Sample/h (1 μs to 1 hr.) |
Pulse mode | 10 k Samples/s to 1 Sample/h (100 μs to 1 hr.) | |
Built-in RAM | 2 million samples for each channels | |
Measurement mode | Logic input mode or Pulse input mode (*2) | |
Pulse input mode | Rotation count (RPM), Accumulating count, Instant count | |
Rotation count (RPM) | Function | Counting the number of pulses per sampling interval and then it is converted to RPM |
Span | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M rpm Full Scale | |
Accumulating count | Function | Accumulating the number of pulses from the start of measurement |
Span | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M counts Full Scale | |
Instant count | Function | Counting the number of pulses per sampling interval (count is reset at each sampling) |
Span | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M counts Full Scale | |
Maximum input frequency | 1 MHz | |
Maximum number of count | 15 M counts (24 bits counter is used) | |
Input signal | Voltage range | 0 to 24 V (common ground) |
Signal type | Contact (Relay), Open collector, Voltage | |
Threshold | Approx. 2.5 V | |
Hysteresis | Approx. 0.5 V (2.5 V to 3 V) | |
Filter | Off or On (-3 dB at 50 Hz) | |
External dimensions (W×D×H) | Approx. 49 x 136 x 160 mm (Excluding projections) | |
Weight | Approx. 700 g |
- *1
- Input prove (RIC-10) is required to connect signals.
- *2
-
The measuring mode is set in each module (16 channels). In Logic mode, up to 7 modules
(Up to 112ch.) can be attached to one main module. In Pulse mode, up to 2 modules
(Up to 32ch.) can be attached to one main modules. The maximum number of module
and channels are limited to up to 10 units with a mixed condition and 112 channels.