Obsnap Instruments Sdn Bhd
29A, Jalan SS 15/4C, 47500 Subang Jaya, Selangor, Malaysia.
+603-5621 5786
+603-5621 5829
+60122185651 (Enquiry)
+60162233687 (Vacancy)
Datalogger Multimeter TES-2732A
Previous | 2 / 7 | Next |
Product Features
- Easy Data Acquisition (Field/Laboratory)
- DMM talks with P.C. via RS-232 interface
- Data Logger function for Field recording without P.C. can record up to 4048 data
- Digital/Analog/Graphic/Listing readout
- Printer to print out DATA
- Direct input of V, A, Ohm, Hz, Cx & ℃/℉ without any further signal modification
- Go/No Go (alarm output) used for control purpose
ELECTRICAL SPECIFICATIONS :
Accuracies are ±(...% of reading + ...digits) at 23℃±5℃,below 80% RH.
- DC Voltage
Range | Resolution | Accuracy | Input Impedance | Overload Protection |
200mV | 0.1mV | 0.5%+1 | 10MΩ | 600Vrms |
2V | 1mV | 1100Vrms | ||
20V | 10mV | |||
200V | 100mV | |||
1000V | 1V | 0.5%+2 |
- AC Voltage ( 50Hz to 500Hz )
Range | Resolution | Accuracy | Input Impedance | Overload Protection |
200mV | 0.1mV | 1%+5 | 10MΩ | 600Vrms |
2V | 1mV | 750Vrms | ||
20V | 10mV | |||
200V | 100mV | |||
750V | 1V | 1.2%+5 |
- DC Current
Range | Resolution | Accuracy | Burden Voltage | Overload Protection |
200uA | 0.1uA | 1%+1 | 0.35V |
0.5A / 250V Fast Blow Fuse & Diode |
2mA | 1uA | |||
20mA | 10uA | |||
200mA | 100uA | |||
20A | 10mA | 1.2%+3 | 0.8V |
20A / 380V Fast Blow Fuse 20A for 30sec max. 10A continuity |
- AC Current ( 50Hz to 500Hz )
Range | Resolution | Accuracy | Burden Voltage | Overload Protection |
200uA | 0.1uA | 1.2%+3 | 0.35V |
0.5A / 250V Fast Blow Fuse & Diode |
2mA | 1uA | |||
20mA | 10uA | |||
200mA | 100uA | |||
20A | 10mA | 1.5%+5 | 0.8V |
20A / 380V Fast Blow Fuse 20A for 30sec max. 10A continuity |
- Resistance ( Ω )
Range | Resolution | Accuracy | Max. Open Circuit Voltage | Overload Protection |
200Ω | 0.1Ω | 0.8%+2 | 3.2V | 600Vrms |
2KΩ | 1Ω | 0.5V | ||
20KΩ | 10Ω | |||
200KΩ | 100Ω | |||
2MΩ | 1KΩ | |||
20MΩ | 10KΩ | 1.5%+3 |
- Diode ( )
Range | Resolution | Accuracy | Max. Open Circuit Voltage | Max. Test Current | Overload Protection |
1mV | 2%+ 2 | 3.2V | 1.0mA | 600Vrms |
- Continuity ( )
Range | Continuity Threshold | Max. Open Circuit Voltage | Overload Protection |
Continuity audible tone for tested resistance below 30Ω | 3.2V | 600Vrms |
- Frequency ( Hz ) ( Test Range 10Hz - 10MHz Auto Range )
Range | Resolution | Accuracy | Sensitivity | Overload Protection |
2KHz | 1Hz | 0.5%+2 | 1.5Vrms | 600Vrms |
20KHz | 10Hz | |||
200KHz | 100Hz | |||
2MHz | 1KHz | |||
10MHz | 10KHz | 3Vrms |
- Temperature Probe used : K (CA) type sensor
Range | Resolution | Accuracy |
℃ | 0.1/1℃ |
0℃~ 200℃ ±(0.5%+1.5℃ ) -50℃~ 0℃ ±( 1% + 2℃ ) 200℃~ 1300℃ ±( 1% + 2℃ ) |
℉ | 0.1/1℉ |
32℉~ 200℉ ±(0.5% + 3℉ ) -58℉~ 32℉ ±( 1% + 5℉ ) 200℉~ 1999℉ ±( 1% + 4℉ ) |
- Capacitance (F)
Range | Resolution | Accuracy | Test Frequency |
2000pF | 1pF | 3%+10 | 400Hz |
200nF | 100pF | ||
2uF | 1nF | ||
20uF | 10nF |
3 1/2 Digital LCD Display
Maximum Reading 1999.
Sampling Rate: 2.5 readings per second
Operating/Storage Condition:
0℃ to 40℃<80% R.H.
-10℃ to 60℃<70% R.H.
Power Source: one 9V Battery
Battery Life Approx. 60hours (Alkaline Battery)
9V Power adaptor jack for long time application
Size & Weight: 180(L)×82(W)×38(H)mm & 365g
Accessories: Test leads ( pair), Spare fuse (0.5A/ 250V ), Battery, Instruction manual, Software, RS-232 Cable, “K” type thermocouple (TP-K01).