Obsnap Instruments Sdn Bhd
29A, Jalan SS 15/4C, 47500 Subang Jaya, Selangor, Malaysia.
+603-5621 5786
+603-5621 5829
+60122185651 (Enquiry)
+60162233687 (Vacancy)
Ultrapure Water Systems
Previous | 2 / 2 | Next |
The PURIST® Ultrapure Water System produces ultrapure water from EDI, reverse osmosis (RO), distilled or deionized water.
Specifications
Features£º
- Automatic system recirculation every hour to prevent bacteria from growing inside the system
- A dual UV lamp to kill bacteria and reduce organics to trace level
- Volumetric dispensing and calibration
- Alarm and maintenance info on a large color LCD monitor
- Adaptable to various final filters. (e.g., final filters for HPLC or to remove RNAse/ Endotoxins)
- Real-time Resistivity measurement by a high-precision resistivity meter with temperature compensation (conductivity cell constant 0.01cm-1)
- Small compact footprint that occupies minimal space in your lab
- Ergonomically designed to ensure a simple installation, operation and easy maintenance
- Manufacturing plant is ISO9001:2008 certified
- Preparation/dilution of HPLC mobile phases, blanks, sample and buffers in GC, HPLC, AA, ICP-MS and other advanced analytical techniques.
- Preparation of media for mammalian cell culture, IVF, reagents for molecular biology applications.
Specifications
Feed Water Quality | |
---|---|
Feed Water Type |
Reverse Osmosis (RO) water Distilled water Deionized or EDI water |
Operating Temperature | 5¡«45 °C |
Feed Water Pressure | 0-15 psi (0-1 bar) |
Product Water Flow Rate | 1.0 - 1.5 L/min at 25°C |
Product Water Quality | |
Resistivity | 18.2 MΩ•cm at 25°C |
TOC | <10 ppb (<5ppb with a dual wavelength UV lamp) |
Particle (>0.2 µm) | No particles (with a 0.2 µm final filter) |
Microorganisms | <0.01 cfu/ml £¨with a 0.2 µm final filter£© |
Pyrogens (endotoxins) | <0.001Eu/ml (with a terminal ultrafiltration cartridge) |
Dimensions | |
Length x Depth x Height | 21 cm× 36cm×44 cm |
Net Weight (Operating Weight) | 6.0 kg (10.5 kg) |